Non-thermal internal energy distribution of ions observed in an electrospray source interfaced with a sector mass spectrometer
作者:David Rondeau、Nicolas Galland、Emilie-Laure Zins、Claude Pepe、László Drahos、Károly Vékey
DOI:10.1002/jms.1873
日期:2011.2
The internal energy distribution P(Eint) of ions emitted in an electrospray (ESI) source interfaced with a sector mass spectrometer is evaluated by using the experimental survival yield (SY) method including the kinetic shift. This method is based on the relationship between the degree of fragmentation of an ion and its amount of internal energy and uses benzylpyridinium cations due to their simple
通过使用包括动力学位移在内的实验存活率(SY)方法,评估与扇区质谱仪对接的电喷雾(ESI)源中发射的离子的内部能量分布P(E int)。该方法基于离子的裂解程度与其内部能量之间的关系,并由于其简单的裂解方案而使用了苄基吡啶鎓阳离子。进行量子化学计算,即在理论的G3(MP2)// B3LYP和QCISD / MP2级别进行。结果表明,与扇区分析仪相连的ESI源中发射的离子的内部能量分布非常狭窄。MassKinetics软件用于确认这些观察结果。在P(ēint)是使每个取代的苄基吡啶鎓阳离子的实验SY与MassKinetics软件生成的理论质谱相符的参数。产生的内部能量分布类似于通过实验SY方法获得的内部能量分布。这表明在目前的实验条件下,P(E int)不能与“热似” Boltzmann分布进行比较。此外,似乎使用扇区分析器在ESI源的第一个泵送阶段增加碰撞能量并不意味着所产生离子的预热。版权所有©2011